PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model P53 pH/ORP Analyzer (universal-mount 1/2 DIN)
Rev. 6-501
48
SET FILTER Time
Select SCALE 0 mA/
4 mA (low endpoint)
A time constant (in seconds) can be set to filter or “smooth
out” the output signal. A minimum value of “0 seconds” has
no smoothing effect. A maximum value of “60 seconds” pro-
vides maximum smoothing. Deciding what output filter time
to use is a compromise. The higher the filter time, the longer
the output signal response time will be to a change in the
measured value.
1.
With the
6(7 P$ 9$/8(
6(7 P$ 9$/8(
6(7 75$16)(5
6(7 ),/7(5
6&$/( P$P$
(;,7
screen displayed, use
Ø
Ø
key
to select the “SET FILTER” line.
2.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust the dis-
played value to the desired filter time, and press
ENTER key
to enter it.
Each output can be set to be 0-20 mA or 4-20 mA.
1.
With the
6(7 P$ 9$/8(
6(7 P$ 9$/8(
6(7 75$16)(5
6(7 ),/7(5
6&$/( P$P$
(;,7
screen displayed, use
Ø
Ø
key
to select the “SCALE 0mA/4mA” line.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view both
choices. With the desired choice displayed, press
ENTER key
to enter this selection.