PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 6-501
Model P53 pH/ORP Analyzer (universal-mount 1/2 DIN)
51
SET TRANSFER Mode
(relay on or off)
From the moment output hold is initiated (during cali-
bration, from TEST/MAINT menu or by TTL input), the
elapsed time INTERVAL or DURATION countdown for
a TIMER relay is temporarily suspended. Also, any
TIMER relay counting down DURATION time is turned
off. When output hold is released, a TIMER relay re-
sumes its INTERVAL or DURATION countdown from
the suspended time. When a TIMER relay is counting
down DURATION time, both outputs are temporarily
held until after the preset DURATION time (and OFF
TIME, if used) elapses.
1.
With the
6(7 )81&7,21
6(7 75$16)(5
$&7,9$7,21
(;,7
screen displayed, use
Ø
Ø
key
to select the “SET FUNCTION” line.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view the
choices (ALARM, CONTROL, STATUS or TIMER).
With the desired choice displayed, press
ENTER key
to
enter this selection.
Normally, each CONTROL and ALARM relay is active, re-
sponding to the measured value of its assigned parameter
(pH or ORP, or temperature). During calibration, however,
you can transfer (XFER) each relay to a preset on/off
transfer state to suit your application requirements:
1.
With the
6(7 )81&7,21
6(7 75$16)(5
$&7,9$7,21
(;,7
screen displayed, use
Ø
Ø
key
to select the “SET TRANSFER” line.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view both
choices (DE-ENERGIZED or ENERGIZED). With the
desired choice displayed, press
ENTER key
to enter
this selection.