PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model P53 pH/ORP Analyzer (universal-mount 1/2 DIN)
Rev. 6-501
42
Select
PULSE SUPPRESS
(on/off)
ENTER NOTE (top line
of MEASURE screen)
2.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust the dis-
played value to the desired filter time, and press
ENTER key
to enter the value.
Sometimes an external interference may occasionally cause
the measurement system to provide unstable readings.
Common causes include entrained gas bubbles in the proc-
ess, and electromagnetic interference (EMI or “electrical
noise” pulses). The analyzer has a pulse suppression fea-
ture to counteract this condition and stabilize readings.
Example: Suppose the analyzer reading is steadily showing
7.3 pH, then suddenly jumps to 9.8 pH for a few seconds,
and returns to 7.3 pH. By turning on this feature, the ana-
lyzer will perceive this as a temporary upset, “suppressing”
most of this pulse change and providing a smoother meas-
urement reading.
1.
With the
',63/$< )250$7
6(/(&7 %8))(5
385( +2 &203
6(7 ,62 32,17
6(7 ),/7(5
38/6( 68335(66
(17(5 127(
7(03 (/(0(17
(;,7
screen displayed, use
Ø
Ø
key
to select the “PULSE SUPPRESS” line.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view both
choices (OFF or ON). With the desired choice dis-
played, press
ENTER key
to enter this selection.
The top line of the MEASURE screen is factory set to read
“PH.” This notation can be changed, for example, to “BASIN
1” to tailor the analyzer MEASURE screen to the applica-
tion. The top line would then be “MEASURE BASIN 1.” The
notation is limited to eight characters which can be a com-
bination of capital letters A through Z, numbers 0 through 9,
and spaces.
1.