PART THREE - OPERATION
SECTION 6 - TEST/MAINTENANCE
Model P53 pH/ORP Analyzer (universal-mount 1/2 DIN)
Rev. 6-501
76
6.8 SIM SENSOR Setting
6.9 RESET CONFIG
Values to Factory
Defaults
•
SENSOR:
Selects the simulated value to be a pH
(or ORP) value.
•
TEMPERATURE:
Selects the simulated value to be
a temperature value.
3.
With the desired choice displayed, press
ENTER key
to
enter this selection and return to the TEST/MAINT top-
level menu screen.
After selecting the type of simulated measurement (subsec-
tion 6.7), set the desired simulation value.
1.
With the
screen displayed, use
Ø
Ø
key
to select the “SIM SENSOR” line.
2.
Press
ENTER key
to display a screen like
.
NOTE:
Both analog output signals are now active.
They have a mA value that corresponds to the
measurement value shown on this screen. (The
relays, depending on their configured settings,
may also respond to this simulation value.)
3.
Use
arrow keys
to adjust the displayed simulation
value to the desired value.
4.
To remove the simulated output and return to the
TEST/MAINT top-level menu screen, press
ESC key
or
ENTER key
.
You can conveniently reset all stored configuration settings
--
but not calibration settings
-- (all at the same time) to
their factory-set defaults shown in Table B.