GIMA SpA
DIATERMO MB250 – DIATERMO MB400
MA303IGBa-19-03-2007 Manuale
d’Istruzioni / Instruction’s Manual
53
9.5 REPAIRS
High frequency cables and electrode holder handle cannot be repaired. Always substitute a damaged part with a new one.
9.5.1
F
USE
S
UBSTITUTION
Before substituting the fuse, disconnect the unit from the mains system.
For substituting use
fuse of the kind 2xT5A (slow) (230V
~
mains voltage) or 2xT10A (115V
~
mains voltage) for
DIATERMO MB400
, 2xT4A (slow) (230V
~
mains voltage) or 2xT8A (115V
~
mains voltage) for
DIATERMO MB250
,
and proceed as it described in chapter INSTALLATION.
9.6 CHECKING OF THE EQUIPMENT BEFORE EACH USE
Each time the use of the electrosurgical equipment is planned a check of the most important safety aspects has to be
implemented considering at least the following:
•
Check the integrity of cords, connections, wires breakage, etc.
•
Assure that all the electrical equipment is properly grounded
•
Assure that all the accessories that should be used are available and sterilised.
•
Inspect the integrity of the packaging of the sterile products.
•
Check, by disconnecting the reference electrode cable, the functioning of the OC light and sound alarm warning.
•
Check, by activating the CUT and COAG power switch, the functioning of the emission lights and sounds warnings.
9.7 FUNCTION AND SAFETY CHECK AND TEST
At least once a year, the biomedical engineering department or other qualified personnel should do the following check and
test:
•
Check of the connectors and mains supply cord conditions;
•
Visual check of the mechanical protections;
•
Check of the protections against the danger due to liquid's pouring, dripping, moisture, liquid's penetration, cleanliness,
sterilization and disinfection.
•
Check of the Equipment's Data on the Label
•
Check of the availability of the Instruction's Manual
•
Check the functioning of the H.F. output controls
•
Check the uniformity of the resistance through the surface of the patient plate.
•
Test the earth conductivity resistance.
•
Test the earth leakage current.
•
Test H.F. leakage current.
•
Control of the neuromuscular stimulation.
•
Control of the accuracy of the output power.