GIMA SpA
DIATERMO MB250 – DIATERMO MB400
MA303IGBa-19-03-2007 Manuale
d’Istruzioni / Instruction’s Manual
40
6.
INSTALLATION
•
Inspect the unit for damages during transport. The claims for possible damages will be accepted only in case they are
immediately communicated to the carrier; the damages that are found must be written down and presented to LED SpA or
to your own retailer. If the unit is returned to the LED SpA or to your own retailer, it is necessary to use the original
equipment’s package or another equivalent one, to guarantee the safety during the transport.
•
Unpack the equipment and carefully study the documentation and operating instruction supplied. Mains voltage, indicated above the
inlet, must agree with the local mains voltage (mains voltage frequency: 50-60 Hz). The equipments predisposed for the main voltage
115/230V
~
are supplied for the main voltage at 230V
~
, in case of supply at 115 V it’s necessary to change fusibles with the value
written on the label.
•
The predisposition of the correct mains voltage is performed in the following way:
(A-B) Extract the fuse holder drawer from the power module.
(C) Insert the fuses making reference to the following chart:
DIATERMO MB400
DIATERMO MB250
Mains Volt.: 110-120 V Delayed fuse 2xT 5A
Delayed fuse 2xT 4A
Mains Volt.: 220-240 V Delayed fuse
2xT 10A
Delayed fuse
2xT 8A
(D) Extract and rotate the detachable part in way to read the correct voltage in the (E) window – reinsert the fuse
holder in the module
•
Connect mains cable to a mains outlet having good hearth connection
OPERATION OF THE EQUIPMENT WITHOUT EARTH CONNECTION IS FORBIDDEN
•
The unit must be installed on a level surface, with dimension, at least, correspondent to those of the base of the unit itself.
Around the unit must be left a space of 25cm, at least.
•
Connect the mains cable to the mains socket on the back panel of the unit.
•
Connect the equipotential binding post located at the left of the unit’s back panel to equipotential socket of the plant.
Connect the double to the connector on the frontal panel
(in the MONOPOLAR 1 section, for monopolar cut and
coagulation;
in the BIPOLAR section
,
for bipolar cut and coagulation) present on the frontal panel of the unit.
•
Connect handle to the corresponding connectors and in the case of use of handle without push-buttons it shall be
connected on the active buckle.
•
Let unit work in dry environment only. Any verified condenses must be let evaporate before putting in operation the unit.
Don't exceed the temperature environment or the allowed moisture.
•
Environmental conditions:
WORK
TRANSIT/STORAGE
Temperature:
from 10 °C up to 40 °C
from -10 °C up to 50 °C
Relative moisture: from 30% up to 75%
from 10 % up to 100%
Atm. Press.: from 70 kPa up to 106 kPa
from 50kPa up to 106kPa
•
When the unit is switched on, through the on/off switch on the frontal panel, after having checked the internal parameters,
it will work with the function and the power level utilized during the last switching (when the unit is switched for the first
time the level will be 00).
•
Before using the unit, it is necessary connect the cable to the patient plate. Both when single plate electrodes and when
split plate electrodes are used it is necessary to confirm the impedance acceptance by pressing the key OK (see paragraph
7
.3.2
). In this way, if the value of the impedance is acceptable, the OC indicator light will stop flashing and the alarm to
sound.
•
The handle for CUT and COAG 1 functions of the MONOPOLAR1 section must be connect to the corresponding output
CUT/COAG1.
•
For the monopolar cut and coagulation (CUT e COAG 2), with the MONOPOLAR 2 currents, connect the handle to the
connector CUT/COAG 2 and the foot-switch to the corresponding output, present in this section.
•
For bipolar cut and coagulation, connect the bipolar forceps and/or the double foot-switch to the respective outputs in the
BIPOLAR section.
The equipment can be connect to external Argon unit by the connector on the back panel (see paragraph 7
.11.1
).