Name
Ambulatory Blood Pressure Monitor
Display
2.4’’ color LCD Display
NIBP Specifications
Measurement Method
Oscillometric method
Working modes
Automatic
Measurement Range
Pressure
adult
0~290mmHg(0~38.6kPa)
pediatric
0~235mmHg(0~31.3kPa)
neonatal
0~140mmHg(0~18.6kPa)
Pulse: 40~240/min
Inflation
adult
160mmHg
pediatric
120mmHg
neonatal
70mmHg
Alarm Range
adult mode
SYS ALM: 40~270 mmHg
DIA ALM: 10~215 mmHg
pediatric mode
SYS ALM: 40~200 mmHg
DIA ALM: 10~150 mmHg
neonatal mode
SYS ALM: 40~135 mmHg
DIA ALM: 10~100 mmHg
Overpressure protect
adult mode
295±5mmHg
pediatric mode
240±5mmHg
neonatal mode
145±5mmHg
Resolution
Pressure
1mmHg
Pulse Rate
±2 bpm or ±2% (select larger)
Measurement Accuracy
Cuff Pressure Accuracy ±3mmHg
Error
The BP Value of the device is equivalence with the
measurement value of Stethoscopy. The error meets all the
conditions in the ANSI/AAMI SP-10:2002+A1:2003 +A2:2006.
Operating
+5ºC~+40ºC
15%RH~80%RH
Temperature/ Humidity
Transport and Storage
-20ºC~+55ºC
≤95%RH
Temperature/Humidity
Atmospheric pressure
80kPa~105kPa
Battery
2 “AA” alkaline batteries
Accessories
Standard Configure:
Adult Cuff: limb circumference 25-35cm (middle of upper arm)
Software CD, User Manual, USB data line, BP extending tube,
A pouch.
Separate Sale:
Pediatric Cuff: limb circumference 10-19cm
(middle of upperarm)
Neonatal Cuff: limb circumference 6-11cm
(middle of upper arm)
SPECIFICATION