L
Leakage
Allows for viewing some real-time measured values for leakage sensor diagnostics. If the leakage
sensor is not connected, the display will show
MODE: SENSOR NOT CONNECTED
and the
voltages VlKR and VlKS will be out of range below.
Leakage Screen
SETTING
DESCRIPTION
VLKR
Voltage at the reference sensor (approx. +0.65 V / +0.95 V).
VLKS
Voltage at the sample sensor (approx. +0.65 V / +0.95 V).
MODE
Mode of the sensor – OFF, STATE FLAG, WARNING, FATAL ERROR.
Service
The partial service menu provides access to the most common operations and is also compatible
with the service menu in older devices.
To access the partial service menu, press the
MenU
key
,
select
SeRViCe (7)
, and then press and
hold the
enTeR
key for three seconds.
Figure 20
Partial Service Menu
The full service menu provides access to all available service operations.
To access the full service menu, press the
MenU
key
,
select
SeRViCe (7)
, and then press the left
arrow, up arrow, right arrow and down arrow buttons in sequence.
Figure 21
Full Service Menu
Содержание VERITY 1741
Страница 1: ...User s Guide VERITY 1741 UV VIS Detector...
Страница 2: ......
Страница 4: ......
Страница 10: ...This page intentionally left blank...
Страница 24: ...This page intentionally left blank...
Страница 28: ...This page intentionally left blank...
Страница 42: ...This page intentionally left blank...