
28
MaInTEnancE
|
VERITY® 1741 UV-VIS DETEcToR
IN
t
e
N
S
It
y C
a
LI
b
r
a
t
IO
N
a
f
t
e
r L
a
M
p(
S
)
r
e
p
L
a
C
e
M
e
N
t
Intensity Calibration after Lamp(s) Replacement
after changing the deuterium lamp, the lamp intensity should be calibrated. If replacing both the
deuterium and tungsten lamps, replace both lamps before performing the intensity calibration.
Follow the instructions below to calibrate the intensity. Refer to
for additional
details.
Instructions
1. Switch on the detector.
2. change the flow cell, if necessary.
NOTE
Calibrate only with the test cell, or with an empty, clean and dry measuring cell. If necessary, you
can use a clean measuring cell filled with non-absorbent liquid (in the whole spectrum).
3. Switch on the lamps.
4. Wait 30 minutes to one hour for the lamps to stabilize.
5. Enter the full service menu by pressing
Menu
|
Service (7)
, and then pressing the left arrow,
up arrow, right arrow, and down arrow buttons in sequence.
6. Select
CaLIbratIONS
and then
INteNSIty
and confirm it.
7. Wait for the calibration to complete (about 12 seconds).
8. check the intensity value (100%) on the initial screen of the display.
9. check for any errors.
10. Remove the test cell and then replace or reinstall the flow cell.
Содержание VERITY 1741
Страница 1: ...User s Guide VERITY 1741 UV VIS Detector...
Страница 2: ......
Страница 4: ......
Страница 10: ...This page intentionally left blank...
Страница 24: ...This page intentionally left blank...
Страница 28: ...This page intentionally left blank...
Страница 42: ...This page intentionally left blank...