Technical Specifications
SPECIFICATION
DEFINITION OR VALUE
Communication Interface
LAN (Ethernet)
Control
Front panel control or computer control via Ethernet and TRILUTION® LC Software
Wavelength Range
200–800 nm (256 CCD elements)
Wavelength Accuracy
± 1 nm
Wavelength Reproducibility
± 0.5 nm
Noise
± 5 x 10
-5
AU using a test cell (254 nm, time constant 1 s)
Refer to ASTM
Drift
1 x 10
-3
AU/hr using a test cell (254 nm after 1 h)
Refer to ASTM
Maximum Data Rate
20 Hz
Flow Cells
Part
Number
Pathlength
Internal
Volume
Maximum
Flow Rate
Connecting Thread
Internal Tubing
14161011
1.3 mm
55 µL
200 mL/min
10-32
1/16" stainless steel tubing
14161014
0.5 mm
50 µL
200 mL/min
10-32
1/16" stainless steel tubing
14161012
0.1 mm
40 µL
200 mL/min
10-32
1/16" stainless steel tubing
14161013
0.05 mm
40 µL
200 mL/min
10-32
1/16" stainless steel tubing
14161015
1.3 mm
55 µL
500 mL/min
¼-28
1/8" FEP tubing
14161018
0.5 mm
50 µL
500 mL/min
¼-28
1/8" FEP tubing
14161016
0.1 mm
40 µL
500 mL/min
¼-28
1/8" FEP tubing
14161017
0.05 mm
40 µL
500 mL/min
¼-28
1/8" FEP tubing
Maximum pressure for all flow cells is 2 MPa (290 psi, 20 bar)
Light Sources
Deuterium and tungsten lamps
Liquid Contact Materials
FEP, fused silica, stainless steel, PEEK, KEL-F
Time Constant
0.1 s, 0.2 s, 0.5 s, 0.75 s, 1 s, 2 s, 4 s, 8 s, 16 s
TECHNICAL SPECIFICATIONS CONTINUED ON PAGE 10
Содержание VERITY 1741
Страница 1: ...User s Guide VERITY 1741 UV VIS Detector...
Страница 2: ......
Страница 4: ......
Страница 10: ...This page intentionally left blank...
Страница 24: ...This page intentionally left blank...
Страница 28: ...This page intentionally left blank...
Страница 42: ...This page intentionally left blank...