Analog Outputs
The detector has four independent analog outputs with cinch connectors. on the rear panel of
the detector they are marked as AnAloG oUT and letters A to D.
Each of the analog outputs can be preset oFFSET in the range of preset values from −0.100 V
to +0.500 V.
A conversion ratio can also be set for output voltage / 1 AU (absorbance unit). This can be set in
the range of 0.2 V / 1 AU to 10.0 V / 1 AU.
In case of a warning about overflow of some of the unused analog outputs, it is possible to turn
off these analog outputs by setting the
Conversion Ratio
to off. Press the
MenU
key and then
select
iO Config | Analog Outputs | AnA 1/2/3/4/ Conversion Ratio
to
OFF
.
Figure 19
Analog Output Cable
ANALOG OUTPUT CABLE MARKING
COLOR
SIGNAL
NOTE
BROWN
HI
AOUT (Analog output)
WHITE
LOW
AGND (Analog ground)
COPPER
SHIELD
AGND (Analog ground)
Offset
The offset of the analog output can be set in the range of −0.100 V to +0.500 V (default value
is 0.00 V).
Conversion Ratio
conversion ratio of output voltage to AU can be set in six ranges: 0.2 V, 0.5 V, 1.0 V, 2.0 V, 5.0 V,
or 10.0 V = 1 AU (default value is oFF). option oFF turns off relevant analog output.
Info
Information about the detector, its components and their condition are displayed.
Device
Displays detector name (
NAME
), its serial number (
SN
) and total operating time
(
TIME
) –
HH:MM:SS
(H – hour, M – minute, S – second).
Содержание VERITY 1741
Страница 1: ...User s Guide VERITY 1741 UV VIS Detector...
Страница 2: ......
Страница 4: ......
Страница 10: ...This page intentionally left blank...
Страница 24: ...This page intentionally left blank...
Страница 28: ...This page intentionally left blank...
Страница 42: ...This page intentionally left blank...