21
VERITY® 1741 UV-VIS DETEcToR
| USER’S GUIDE
t
u
r
N
l
a
M
p O
N
Turn Lamp On
To use TRILUTIoN Lc to turn the lamp(s) on:
NOTE
The detector requires a 15–30 minute warm-up period to achieve a stable baseline.
1. create a TRILUTIoN Lc method with the detector in the configuration.
a. Select
Scan
. The software searches for the detector.
b. From the
available Instruments
window, drag and then drop the detector in the
workspace.
2. on the
Control
tab, drag and then drop the
turn lamp On
task into the control workspace.
3. Save the method, and then select
run
. The
application run
window appears.
4. on the
application run
window, select the method name from the
Method Configuration
drop-down list and then select .
Manual
Control
icon
Method
Configuration
Figure 14
Application Run Window
5. Select the
Manual Control
icon. The
Manual Control
window appears.
6. Tasks are filtered by instrument name. Drag and then drop the
turn lamp On
task into the
workspace. Select
OK
.
Figure 15
Turn Lamp On Task
run
turn lamp On
task
7. Select
run
on the
Manual Control
window.
The instrument is ready for operation.