diGiTAL OUTpUT SWiTCH (dOUT1 And dOUT2)
Function setting for outputs DoUT 1 and DoUT 2 on external interface:
SETTING OF ACTIONS ON DIGITAL OUTPUT
SETTING
DESCRIPTION
OFF
No function.
RTM
The output is possible to control by commands from PC.
OVRs1
Level H, when the absorbance A is higher than THRESHOLD s1.
OVRs2
Level H, when the absorbance A is higher than THRESHOLD s2.
INs12
Level H, when the absorbance A is in the THRESHOLD s1 and s2 range.
OUTs12
Level H, when the absorbance A is out of the THRESHOLD s1 and s2 range.
DIN1
Level copies the input DIN1.
DIN2
Level copies the input DIN2.
LEAK
Level H, when the leak is detected (LEAKAGE).
FERR
Level H, when a fatal error occurs.
LEDER
Setting for externally connected ERROR LED, all errors.
LEDLP
Setting for externally connected LAMP LED.
If the detector is not in ABS mode and the digital output is controlled by threshold values s1 and
s2, then the digital output is always disconnected.
example
The threshold value s1 is set at 0.1 AU and the digital output at oVR s1. If the absorbance signal is
lower than 0.1 AU, then the digital output is disconnected. If the value is higher, then the digital
output is connected. With this type of connection, the output signal can be used for simple
fraction collection.
THReS S1 And THReS S2
Set the threshold value s1 and s2 in the range of −100 to +4000 mAU.
Содержание VERITY 1741
Страница 1: ...User s Guide VERITY 1741 UV VIS Detector...
Страница 2: ......
Страница 4: ......
Страница 10: ...This page intentionally left blank...
Страница 24: ...This page intentionally left blank...
Страница 28: ...This page intentionally left blank...
Страница 42: ...This page intentionally left blank...