VERITY® 1741 UV-VIS DETEcToR
| USER’S GUIDE
chapter 5
Chapter 5 |
trOUBLeShOOtING
Troubleshooting
PROBLEM
CAUSED BY / CONDITIONS
SOLUTION
After switching the detector on, no
LED on the front panel is lit
Line fuse failure.
Check the main fuse on the
rear of the detector see
Electronics failure.
Contact your local Gilson
representative.
Noisy or drifting baseline
Leaking or dirty flow cell, bubbles,
or non-transparent mobile phase.
Check cell connections. Check
mobile phase. Check cell purity.
Electronics/optic failure.
Contact your local Gilson
representative.
Warning or error message
-
Try resetting of error and if is not
successful contact your local Gilson
representative.
Leaking flow cell
Clogged tubing.
Check and replace flow cell inlet/
outlet tubing, if necessary.
Damaged or worn flow cell.
Replace the flow cell.