Chapter 5 Cleaning, Disinfection and Sterilization
- 40 -
cup.
——
One-off sterile cloth.
——
Sterile water.
c) PRODUCTS
——
Cleanser: Bianex® Special Efficiency Multi-Enzymatic Cleanser
——
Disinfectant: Sporicidin Sterilizing and Disinfecting Solution (K003087)
5.8.2 Thorough Cleaning
a) Prepare the cleaning solution by adding 1-2ml Bianex® Special Efficiency Multi-Enzymatic
Cleanser into 500ml of sterile water. Additional amounts maybe required for hard-to-remove
matter. The water temperature should be from 20
℃
to 64
℃
.
b) Immerse the A-Probe and/or eye cup that need to be disinfected into the container with
500ml cleaning solution for 5 minutes (Do not immerse the probe connector). Clean the
A-Probe and/or eye cup carefully in the cleaning solution with a soft silk brush and shake it to
wash; and then take it out and wipe off the cleaning solution with sterile cloth.
c) Rinse the cleaned A-Probe and/or eye cup by immersing them in 500ml sterile water for 1
minute.
d) Drain the A-Probe and/or eye cup and wipe it with sterile cloth.
Note: 1) Disposable gloves, eye protection and liquid-proof gowns should be worn
when cleaning soiled probe and/or eye up.
2) All detail information for Bianex® Special Efficiency Multi-Enzymatic
Cleanser, please refer to the manufacturer’s instruction.
5.8.3 Disinfection
Use disinfectant solution immersion method:
a) Activate and prepare the disinfectant solution (Sporicidin Sterilizing and Disinfecting
Solution) according to the manufacturer’s instructions.
b) Sporicidin Indicators should be used before each use of this solution to determine whether
glutaraldehyde is at or above its minimum concentration of 0.6% and phenol is at or above its
minimum recommended concentration of 1.3%. Do not use beyond 14 days even if the
respective concentration of active ingredients is above the minimum recommended
concentration as indicated by the indicator strips.
c) Disinfect the A-Probe and/or eyecup by immersing them completely in about 500ml of
disinfectant solution for a minimum of 20 minutes at 25
℃
. Please keep the probe connector
dry during disinfection.
d) Drain it and wipe off the disinfectant solution with sterile cloth.
Содержание GRU-5000 A
Страница 1: ......
Страница 61: ...Annex D Guidance and Manufacturer s Declaration 52 Annex D Guidance and Manufacturer s Declaration...
Страница 62: ...Annex D Guidance and Manufacturer s Declaration 53...
Страница 63: ...Annex D Guidance and Manufacturer s Declaration 54...
Страница 64: ...Annex D Guidance and Manufacturer s Declaration 55...
Страница 68: ......