WindSonic Doc No 1405 PS 0019 Issue 28 November 2019
68
Commands and Returns in Low Power Mode (aXLPM!)
Note:- Rolling average data not available.
Use of the aR2!, aR3!, aR4! and aR5! (and their associated aRCn!) commands when the
device is in aXLPM! (low power - continuous) mode, will return a "zero data values ready"
message.
Name
Command
Gill Output
Not Applicable
aR0!
Zero Data values ready e.g. 0<CR><LF>
Not Applicable
aRC0!
Zero Data values ready e.g. 0AP@<CR><LF>
Not Applicable
aR1!
Zero Data values ready e.g. 0<CR><LF>
Not Applicable
aRC1!
Zero Data values ready e.g. 0AP@<CR><LF>
Not Applicable
aR2!
Zero Data values ready e.g. 0<CR><LF>
Not Applicable
ARC2!
Zero Data values ready e.g. 0AP@<CR><LF>
Not Applicable
AR3!
Zero Data values ready e.g. 0<CR><LF>
Not Applicable
ARC3!
Zero Data values ready e.g. 0AP@<CR><LF>
Not Applicable
AR4!
Zero Data values ready e.g. 0<CR><LF>
Not Applicable
ARC4!
Zero Data values ready e.g. 0AP@<CR><LF>
Not Applicable
AR5!
Zero Data values ready e.g. 0<CR><LF>
Not Applicable
ARC5!
Zero Data values ready e.g. 0AP@<CR><LF>
Start Measurement (Polar)
aM!
atttn<CR><LF> e.g. a0053
Where a is unit identifier, 005 is measurement time in
seconds and 3 is the number of data readings.
Send Data
aD0!
a<dir><mag><status><CR><LF>
e.g. 0+083+000.02+00<CR><LF>
Start Measurement (UV)
aM1!
atttn<CR><LF> e.g. 00053
Send Data
aD0!
a<U><V><status><CR><LF>
e.g. 0+000.00+000.02+00<CR><LF>
Start Measurement with
CRC (Polar)
aMC!
atttn<CR><LF> e.g. 00053
Send Data
aD0!
a<dir><mag><status><CRC><CR><LF>
e.g. 0+295+000.01+00B{T<CR><LF>
Start Measurement with
CRC (UV)
aMC1!
atttn<CR><LF> e.g. 00053
Send Data
aD0!
a<U><V><status><CRC><CR><LF>
e.g. 0+000.00+000.00+00JZY<CR><LF>
Acknowledge Active
a!
a<CR>< LF> e.g. 0<CR><LF>
Send Identification
aI!
a13GillInst 2487 B06 <sn><CR><LF>
e.g. 013GillInst 2487B0611220012<CR><LF>
Change Address
aAb!
b<CR><LF> e.g. 1<CR><LF>
Address Query
?!
a<CR><LF> e.g. 0<CR><LF>
Measurement on demand
aXLPM!
0LP<CR><LF>
Continuous Reading
aXHPM!
0HP<CR><LF>