Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
MetPak/MetPak RG and MetPak Pro Base Stations
Page | 17
Issue 1
Doc. No. 1723-PS-0023
October 2018
Whilst squeezing the red retainer in the direction of ARROWS A, pull in the direction of ARROW B.
A
A
B
Your connector should now resemble the connector in the picture below.
Insert each contact pin until you feel a slight click. If you have inserted the contact into the
incorrect hole it can be removed at this point by simply pulling it out. Please note there will be
some resistance.
WindSonic (Option 2/3/M) and WindMaster/Pro RS422 10 Metre Cable Connections are as
follows:-
Base Station 10 Metre Cable
WindSonic Option 2
and 3 Connector Pin
Wind Sonic M
Non Heated
WindMaster
/WindMaster Pro
Green White (Sig Gnd)
1
No Connection
4
Orange White ()
2
2
11
White Orange (Supply -)
3
3
12
Brown White RS422+ input)
4
4
2
White Brown (RS422– input)
5
5
1
Blue White (RS422+ output)
6
6
5
White Blue (RS422- output)
7
7
6
White Green (Screen shield)
Cut off wire at sleeve
1
31
Rear View of WindSonic Connector
Rear View of WindMaster Connector