Operation
FSH-2-PH
4.02.5
E.A.C. Collector Cell / HEPA Filter Installation
4.02.4
E.A.C. Collector Cell / HEPA Filter Removal
NOTE: HEPA
Filter models ONLY
‐
charcoal filter is located beneath the HEPA, such that air passes through it first before the
HEPA
filter. When installed, the filter must properly engage a proofing switch located on the back wall.
EAC
EAC
①
Unlatch
Access Cover
and lift off.
②
Grasp
E.A.C.
Collector Cell / HEPA
Filter
and pull straight out, at a slightly upward angle.
1
2
EAC
Pin must align & fit
thru hole in rear of
cell.
Airflow
arrow on
cell must
point up
①
Ensure the airflow indicator
arrow points
[UP]
& the
cell’s
Contact Pad
faces to
the
right side
.
②
Align cell (or filter) in
guides
and slide into hood. If
installed properly, cell is
flush with hood front.
If
not, cell is not installed
incorrectly
.
③
Replace
Access Cover
and latch.
1
2
EAC
Содержание FSH-2-PH
Страница 4: ......
Страница 14: ...4 Introduction FSH 2 PH ...
Страница 36: ...26 Overview FSH 2 PH ...
Страница 45: ...35 Operation FSH 2 PH NOTES ...
Страница 53: ...43 Cleaning Maintenance FSH 2 PH ...
Страница 56: ...46 Troubleshooting FSH 2 PH ...
Страница 58: ...48 Parts List FSH 2 PH 1 2 3 5 6 8 7 4 Not shown 7 02 Ventless Hood Front ...
Страница 60: ...50 Parts List FSH 2 PH 4 7 5 6 1 2 3 8 9 10 11 12 13 14 15 ILS Only Not Shown 7 03 Control Box ...
Страница 62: ...52 Parts List FSH 2 PH 3 4 1 2 6 10 7 9 11 13 12 5 8 17 14 16 18 19 20 21 22 23 24 15 7 04 Ventless Hood Rear ...
Страница 64: ...54 Parts List FSH 2 PH Notes ...
Страница 65: ......