Special Modes
Giga-tronics 2500B Series Microwave Signal Generators
Page 68
Operation Manual, Part Number 34737, Rev C, June 30, 2011
6.1
Use Cable Correction
Cable Correction allows you to adjust the output power of the 2500B to compensate for losses created by connecting
devices such as RF cables, connectors and other RF devices. The maximum correction is less than 10 dB, and depends on
the absolute power output level. The power correction cannot exceed the maximum or minimum power output within a
given setting of the step attenuator.
This function creates a user correction table of 1001 points for the frequency range of the unit. The power output for
frequencies between the correction points is determined using a linear interpolation algorithm for the first frequency
points above and below the selected frequency.
NOTE:
The example procedure in this section uses specific parameters to illustrate how to setup the 2500B. You can use
this procedure for real-life situations by simply changing the parameters to fit your needs.
Required Equipment
Model 2500B series Microwave Signal Generator with firmware version 4.09 or higher
Giga-tronics 8650A or 8650B series Universal Power Meter
Giga-tronics 803XXA series CW Power Sensor, or other sensor appropriate to the frequency range and signal
type to be calibrated (with power sensor cable)
BNC Male to Male Coaxial Cable
RS-232 Null Modem Cable, Male to Male (DB9)