Giga-tronics 2500B Series Microwave Signal Generators
Performance Verification
Operation Manual, Part Number 34737, Rev C, June 30, 2011
Page 123
9.2.13
Pulse Modulation: Rise and Fall Time
This test measures the rise and fall times of pulse modulation.
9.2.13.1
Equipment and Materials
Oscilloscope, > 500 MHz bandwidth recommended
Pulse generator
Crystal (diode) detector, < 10 ns rise time, ≤ 3 pF output capacitance , frequency range equal or greater than the
test frequency range
Cables and adapters for connecting the test setup
Datasheet 10 (Table 80 on page 125)
Figure 54: Pulse Modulation Rise and Fall Time Test