CW Operation
Giga-tronics 2500B Series Microwave Signal Generators
Page 26
Operation Manual, Part Number 34737, Rev C, June 30, 2011
Generate a CW Signal
Step
Action
9.
To compensate for a loss that varies linearly with frequency
;
1.
Select
Power Slope.
2.
Enter the desired correction factor using the numeric keypad,
keys, or
. Note that when
this correction factor is entered,
SLP
appears in the upper right corner of the display.
10.
To adjust the phase of the output
; select
Phase Adjust, and enter the desired phase
shift using the numeric keypad,
keys, or
.
11.
If the RF ON indicator is not lit, press the RF ON button to enable the RF output.
12.
Verify that the Unleveled indicator is not lit.
13.
If the Unleveled indicator is lit, then the combination of output power level, power offset, power
slope, and step attenuator mode (if applicable) is set inappropriately, and the RF output is unleveled.
Adjust the combination of settings until the Unleveled indicator turns off.
End of Procedure
Softkey
Softkey