HD50CR
- 32 -
V1.0
F412: lower alarm threshold enabling {On/OFF, default=On}
F413: upper alarm threshold
F414: upper alarm threshold enabling {On/OFF, default=On}
F415: alarm hysteresis in % of the difference between the two alarm thresholds
{0…100 %, default=2%}
F416: delay time in signaling the alarm {max. 960 min, default=0}
F417: buzzer enabling {On/OFF, default=OFF}
F418: buzzer “Latch” function enabling {On/OFF, default=OFF}
F419: buzzer silencing time, in minutes {0…1080, default=0}
.
.
.
F460: alarm of the quantity of index 6
F461: lower alarm threshold
F462: lower alarm threshold enabling {On/OFF, default=On}
F463: upper alarm threshold
F464: upper alarm threshold enabling {On/OFF, default=On}
F465: alarm hysteresis in % of the difference between the two alarm thresholds
{0…100 %, default=2%}
F466: delay time in signaling the alarm {max. 960 min, default=0}
F467: buzzer enabling {On/OFF, default=OFF}
F468: buzzer “Latch” function enabling {On/OFF, default=OFF}
F469: buzzer silencing time, in minutes {0…1080, default=0}
F500 submenu (UNIT MEAS MENU)
Units of measurement.
F501: temperature unit of measurement {°C/°F, default=°C}
F502: differential pressure unit of measurement {hPa / kPa / atm / mmHg / mmH
2
O /
inHg / inH
2
O / (kgf /cm
2
) / PSI / mbar / bar / Pa, default=hPa}
F600 submenu (LOG MENU)
Logging settings.
F601: logging enabling/disabling {On/OFF, default=On}
F602: choice between cyclical management (the new data overwrite the old ones when
the memory is full) or non-cyclical management (logging stops when the
memory is full) of the instrument internal memory {YES=cyclical management
/ NO=non-cyclical management, default=YES}
F603: logging interval (if it is higher than the measuring interval, the average of the
measurements acquired during the interval will be stored) {default=30 s}
F604: measurements acquisition interval {default=1 s}. If a value higher than the
logging interval is set, the measurements will be detected with interval
equal to the logging interval.
F605: deletion of all stored measurements from the instrument memory {YES/nO}
Содержание Delta OHM HD50CR
Страница 44: ...HD50CR 44 V1 0 Dimensions mm...
Страница 49: ...HD50CR 49 V1 0 NOTES...
Страница 50: ...HD50CR 50 V1 0 NOTES...
Страница 51: ......