HD50CR
- 31 -
V1.0
F100 submenu (DEV INFO)
Instrument general information.
F101: model
F102: serial number
F103: user code
F104: group name
F105: firmware version
F106: firmware date
F107: factory calibration date
F108: user calibration date
F109: type of calibration in use (factory or user)
F110: external probe serial number
F200 submenu (MEAS MENU)
Selection and sorting of the quantities to be detected.
F201: temperature measurement index {CH1…CH6, default=CH2}
F202: relative humidity measurement index {CH1…CH6, default=CH3}
F203: Dew Point measurement index {CH1…CH6, default=CH4}
F204: Absolute humidity measurement index {CH1…CH6, default=CH5}
F205: Wet bulb temperature measurement index {CH1…CH6, default=CH6}
F206: differential pressure measurement index {CH1…CH6, default=CH1}
Set a measurement to OFF to exclude it.
F300 submenu (DISP LOOP VIEW)
Selection of the quantities to be displayed cyclically.
Note: in the F300 submenu, only the quantities to be detected appear, in the order set
in the F200 submenu.
F301: quantity of index 1 {On/OFF, default=On}
F302: quantity of index 2 {On/OFF, default=On}
F303: quantity of index 3 {On/OFF, default=On}
F304: quantity of index 4 {On/OFF, default=On}
F305: quantity of index 5 {On/OFF, default=On}
F306: quantity of index 6 {On/OFF, default=On}
Set a measurement to OFF to exclude it from the cyclical display.
F400 submenu (ALRM MENU)
Setting of the alarms.
Note: in the F400 submenu, the quantities appear in the order set in the F200 sub-
menu.
F408: period, in seconds, of the beep emitted by the buzzer {1…3600, default=2}
F409: duration, in seconds, of the beep emitted by the buzzer {0.1…20, default=1.0}
F410: alarm of the quantity of index 1
F411: lower alarm threshold