15

Signet 8150 Flow Totalizer

Speed

SPEED averaging serves to smooth out fluctuations in the flow 

rate that may be caused by inadequate straight pipe runs after 

pumps, valves, and elbows in the pipe. The selections are 0, 7, 

15, 30, 60 and 120 seconds. The factory standard setting is  

30 seconds.

• 

Use faster (0 s - 30 s) averaging for well-established, stable 

flow conditions.

• 

Use slower (60 s - 120 s) averaging if the flow conditions are 

unstable.

Note:  

While the SPEED setting helps to smooth out the 

fluctuations caused by piping conditions, it also causes 

a delay in showing actual changes in flow rate. The 

SENSITIVITY function (section 15) is designed to help 

offset this effect.

 

Example:  Change the SPEED setting from 60 seconds  

to 30 seconds.

1.  Hold the ENTER key for 2 seconds, then enter the 

SECURITY CODE.  The display shows the first item in the 

CALIBRATE menu, AutoCAL.

2.  Press the ▼ key five times to scroll to DISP SPEED.   

The display shows DISP SPEED, the "sec" annunciator, and 

the current speed setting.

3.  Press the ► key to select the Display speed for editing.   

The current speed selection begins flashing.

4.  Press the ▼ key one time to scroll from 60 seconds to 30 

seconds.

5.  Press the ENTER key to return to the menu.

6.  Press ▲ and ▼ keys together to exit the menu and return 

to normal operation.  The display shows "Storing" for a few 

seconds, then returns to the normal operating display.

2 s

1

3

4

5

6

2

5x

Enter Security code

*

**

Pages 16-18 provide K-Factors for the Signet 515 and 525 flow 

sensors. Use this table to convert the K-Factor to other units of 

measure, and to set the Total K-Factor.

NOTE:

The maximum K-Factor is 999999.

The minimum K-Factor is 0.001.

Flow and Total K-Factor Selection

If you 

want the 

FLOW RATE 

to read in:

and you want 

the TOTALIZER 

to count in:

Set the Flow 

K-Factor to:

and set the 

Total 

K-Factor to:

Liters

Liters

K(Liter)

1

Liters

Kiloliters

K(Liter)

1000

Liters

cubic meters

K(Liter)

1000

cubic meters

cubic meters

K(Liter) x 1000

1

cubic meters

Megaliters

K(Liter) x 1000

1000

Kiloliters

Kiloliters

K(Liter) x 1000

1

Kiloliters

Megaliters

K(Liter) x 1000

1000

Megaliters

Megaliters

K(Liter) x 1 000 

000

1

U.S. gallons

U.S. gallons

K(gal)

1

U.S. gallons

U.S. gallons  

x 1000

K(gal)

1000

U.S. gallons

cubic feet

K(gal)

7.4805

U.S. gallons

acre-inches

K(gal)

27154

U.S. gallons

Acre-feet

K(gal)

325848

U.S. gallons

Kiloliters

K(gal)

264.2

Acre-Inches

Acre-Inches

K(gal) x 27154

1

Acre-Inches

Acre-feet

K(gal) x 27154

12

Acre feet

Acre-feet

K(gal) x 325848

1

Acre feet

Acre-Inches

K(gal) x 325848

0.083

cubic feet

cubic feet

K(gal) x 7.4805

1

Содержание Signet 8150

Страница 1: ...150 Flow Totalizer is a battery powered instrument capable of providing uninterrupted flow and total volume information for 4 years and beyond The unique features of the 8150 include Easy setup and di...

Страница 2: ...mical poisoning dry out or mishandling e g broken glass damaged membrane freezing and or extreme temperatures Thank you for purchasing the Signet line of Georg Fischer measurement products If you woul...

Страница 3: ...umidity 0 to 95 non condensing Rating NEMA 4X IP65 front panel Shipping Weight 0 5 kg 1 1 lbs Quality Standards CE CUL UL Manufactured under ISO 9001 for Quality ISO 14001 for Environmental Management...

Страница 4: ...ng it so the square tab is close to the conduit ports 3 Insert the mounting screw into the conduit base so the head of the screw presses down on the locking ring when tightened 4 Connect sensor wires...

Страница 5: ...ied in bold type are required for the installation Other parts are shown for reference only Use 2 conductor shielded cable no more than 30 m 100 ft long Remote Field mount on Pipe Flow Totalizer 3 815...

Страница 6: ...d 6 Ext Reset Two 3 6 V lithium thionyl chloride batteries AA size 7400 0011 are installed in the Totalizer NOTE The 8150 will not operate with standard 1 5 V alkaline batteries USE 3 6 V SAFT LS14500...

Страница 7: ...uring normal operation Illustrated No flow for 3 hours Operation Standard Menu Settings ENTER 5 sec Function Factory set Description AUTO CALIBRATION No setting See page 11 for detailed information FL...

Страница 8: ...lement to the last digit 6 Press the key one time to scroll the flashing digit to 1 7 Press the ENTER key to complete the edit process 8 Press the and keys together to exit the Calibrate menu store th...

Страница 9: ...rge volume Total 2 tot2 can be reset only by entering the security code in the calibration menu Total 2 is identified by a flashing display every six seconds Application Use Total 2 for extended measu...

Страница 10: ...lizer value will begin flashing The totalizer will automatically reset to 00000000 in 8 seconds While the display is flashing you can cancel the reset by pressing and keys together 5 Press and keys to...

Страница 11: ...ted K Factor is less than 0 001 or greater than 999999 Verify the flow rate and start the AutoCAL procedure from step 1 8 Press and keys together to store the new value in the memory The display will...

Страница 12: ...o the decimal point 5 Press the key one time to move the decimal point one position to the right 6 Press the ENTER key to return to the menu 7 Press and keys together to exit the menu and return to no...

Страница 13: ...ay will not auto range If the decimal is set to tenths the flow rate display will show tenths up to 999 9 then the auto range will switch to whole numbers 1000 to 9999 If the decimal is set to hundred...

Страница 14: ...bilized but a sharp change in flow rate is not represented for 60 seconds or longer dotted green line With SPEED at 60 seconds and SENSITIVITY set to 6 the flow rate is stabilized while the sudden shi...

Страница 15: ...ss the ENTER key to return to the menu 6 Press and keys together to exit the menu and return to normal operation The display shows Storing for a few seconds then returns to the normal operating displa...

Страница 16: ...4 IR4T012 62 979 16 639 1 1 2 IR4T015 46 688 12 335 2 IR4T020 29 459 7 7832 BRONZE TEES ON SCH 40 PIPE 1 BR4T010 104 54 27 619 1 1 4 BR4T012 62 979 16 639 1 1 2 BR4T015 46 688 12 335 2 BR4T020 29 459...

Страница 17: ...950 DN 40 SFMT015 43 490 11 490 DN 50 SFMT020 25 908 6 8450 DN 65 SFMT025 18 067 4 7732 DN 80 SFMT030 12 357 3 2648 DN 100 SFMT040 8 0599 2 1294 DN 125 SFMT050 4 4312 1 1707 DN 150 SFMT060 3 2271 0 8...

Страница 18: ...ceiving a signal from the flow sensor 1 There is no flow in the pipe 2 Flow sensor is not turning due to blockage or damage 3 Sensor wiring is loose or incorrect In AutoCal the calculated K Factor is...

Страница 19: ...19 Signet 8150 Flow Totalizer Notes...

Страница 20: ...roof rear cover for panel mount totalizer 3 9000 392 159 000 368 Liquid tight connector kit includes 3 connectors 3 9000 392 1 159 000 839 Liquid tight connector NPT 1 connector 3 9000 392 2 159 000 8...

Отзывы: