13
Signet 8150 Flow Totalizer
Select the timebase for the flow rate. The available selections
are seconds, minutes, hours or days.
Example: Change the Timebase from MINUTES
(factory standard) to DAYS
1. Hold the ENTER key for 2 seconds, then enter the
SECURITY CODE. The display shows the first item in the
CALIBRATE menu, AutoCAL.
2. Press the ▼ key three times to scroll to the Timebase.
3. Press the ► key to select the Timebase for editing.
4. Press the ▲ key two time to scroll from MIN to DAY.
5. Press the ENTER key to return to the menu.
6. Press ▲ and ▼ keys together to exit the menu and return
to normal operation. The display shows "Storing" for a
few seconds, then returns to the normal operating display.
Enter
security
code
2 s
1
3
4
5
6
2
2x
3x
*
**
Timebase
Decimal Point for Flow Display
Select the maximum decimal resolution for the flow rate
display. The available selections are hundredths (xx.xx),
tenths (xxx.x) or whole numbers only (xxxx.). The decimal will
auto-range down to this setting.
•
If the decimal is set to whole numbers, the flow rate
display will not auto-range.
•
If the decimal is set to tenths, the flow rate display will
show tenths up to 999.9, then the auto-range will switch to
whole numbers (1000 to 9999.)
•
If the decimal is set to hundredths, the flow rate display
will show hundredths up to 99.99, then tenths from
100.1 to 999.9, then whole numbers to 9999.
Example: Change the maximum decimal display from
hundredths to tenths:
1. Hold the ENTER key for 2 seconds, then enter the
SECURITY CODE.The display shows the first item in the
menu, AutoCAL.
2. Press the ▼ key four times to scroll to the Decimal setting.
The display shows four dashes and the current decimal
setting.
3. Press the ► key to select the decimal for editing.
The decimal point will begin to flash.
4. Press the ▼ key one time to move the flashing decimal
from hundredths to tenths.
5. Press the ENTER key to return to the menu. The decimal
will stop flashing.
6. Press ▲ and ▼ keys together to exit the menu and return
to normal operation.
7. The display shows "Storing" for a few seconds, then
returns to the normal operating display.
Available Display
Selections
Your Flow
rate
will read on
display as:
hundredths (XX.XX)
10.55
10.55
tenths (XXX.X)
10.55
10.6
whole numbers (XXXX.)
10.55
11
2 s
1
3
4
5
6
2
4x
Enter Security code
**
*