UNIT SPECIFICATIONS SHEET
LFGI-6USP LAMINAR FLOW GLOVEBOX / ISOLATOR
SECTION I. PERFORMANCE DATA
The Laminar Flow Glovebox/Isolator provides the ultimate in both product and operator
protection. The LFGI functions as a Glovebox while operating under negative pressure to
meet NIOSH recommendations and as an Isolator under positive pressure to meet USP
<797> regulations. HEPA filtered unidirectional (laminar) air bathes the work area to
protect the product from contamination and removes any particulate generated by sample
manipulation. The operator is provided a complete barrier from materials being
manipulated in the work area. The LFGI Series of barrier isolators meet or exceed ISO
14644-1 Class 5 air quality. Each LFGI, before shipping, undergoes rigorous physical testing
to assure the unit meets performance requirements as validated. It is required that
independent certification be performed before use.
SECTION II. EQUIPMENT SPECIFICATIONS
Construction-
-Overall Dimensions: 90” W (74” W at base) x 32” D x 80” H
-Work Area Dimensions: 70 ½“ W x 24 ½” D x 29” H
-Designed to fit through standard door openings and elevators.
-All stainless steel construction both inside and out with a pharmaceutical grade finish.
-316L Stainless steel work trays slide over one another and are removable to facilitate easy clean
up under the work deck.
-All viewing panels are stainless steel framed Acrylic
-Front viewing panel is top hinged and self supporting for easy access to the work area for
cleaning and equipment ingress.
-All filter diffuser / guard panels are removable for easy cleaning.
-Straight sides and back maximize work area to accommodate the many types and shapes of
pharmacy equipment and compounders.
-All corners in work area, antechamber and work surface are easily reached and cleaned.
Antechamber-
-Sealed two-door airlock (16” W x 13 ½” D x 14 ¼” H) maintains complete environmental
separation between work area and ambient conditions.
-HEPA filtered purge of trapped airlock air eliminates cross contamination between the
work area and the room during both material ingress and egress.
Waste Removal-
-Trash tube is a 4” diameter tube, flanged for waste bag/container connection and provided
with an easily removable seal/stopper to maintain work area containment. (Stainless trash can is
available)
-Sharps tube is a straight 2 ½” diameter tube to facilitate quick drop of the largest syringes and
provided with an easily removable seal/stopper to maintain work area containment.
-Fully adjustable sharps container shelf accommodates most commercially available wide mouth
or hinged top sharps collectors. (Examples BD Part # 305610 and 305491)
Safety-
56
Содержание LFGI-3USP
Страница 1: ...Laminar Flow LFGI CAI or CACI User Manual ...
Страница 33: ...33 ...
Страница 51: ...SECTION IV UNIT DIAGRAM Made in the USA and available for purchase on the GSA 51 ...
Страница 55: ...55 ...
Страница 59: ...59 ...
Страница 63: ...Fig 2 Air recycle plenum connection tip removed 63 ...
Страница 85: ... 863 248 2800 www sensocon com 20 Installation and Operation Manual Series A3 A4 ...