Modification of glove change procedure-
Changing the glove while in operation to maintain containment is possible and can become
easy with practice.
1) Pull glove and sleeve inside-out of the work area.
2) Remove the glove O-ring.
3) Roll the existing glove’s cuff down to the first groove in the GOS (the one closest to your
body).
4) Take a new glove and stuff it inside of the existing glove leaving the cuff out and ensuring
the thumb is pointed up.
5) Stretch the new glove’s cuff over the old one up to the middle groove of the GOS.
6) Secure the new glove with the glove O-ring at the middle groove of the GOS.
7) Push the glove and sleeve back into the work area.
8) Reach into the work area through another glove and grab the new glove assembly. At
this point the glove exposed to the work area is the old glove with the new glove trapped
behind it.
9) With two fingers, carefully work the old glove off of the GOS from inside the work area.
It will begin to come free in one spot. Work the free spot around the perimeter of the
GOS removing the old glove.
10) Dispose of the old glove through the waste tubes and push a hand into the new glove.
Sleeve Changes
As seen in this picture there are two grooves on the
LFGI gloveport. The sleeve can be placed at either
position and secured with the Sleeve O-ring.
To change a sleeve during maintenance, simply
remove the securing O-ring and pull the sleeve out
of the work area and off of the gloveport.
Typically, new sleeves do not come with GOS or
glove O-rings so save them from the existing sleeve.
To order extras contact GERMFREE at 800-888-
5357.
To replace while maintaining containment of the
work area first remove the sleeve O-ring and move
the sleeve to the outermost groove. Next push the
new sleeve inside of the old and pull over the
gloveport down to the innermost groove. Replace
38
Содержание LFGI-3USP
Страница 1: ...Laminar Flow LFGI CAI or CACI User Manual ...
Страница 33: ...33 ...
Страница 51: ...SECTION IV UNIT DIAGRAM Made in the USA and available for purchase on the GSA 51 ...
Страница 55: ...55 ...
Страница 59: ...59 ...
Страница 63: ...Fig 2 Air recycle plenum connection tip removed 63 ...
Страница 85: ... 863 248 2800 www sensocon com 20 Installation and Operation Manual Series A3 A4 ...