Exhaust-
-Clean air plenum is external to the work area.
-Constructed to allow for optional outside venting of exhaust air.
-Constructed to allow for 3 exhaust options that can be factory set or changed by a certifier in-
place
-Venting is not required for operation but is optional.
No Exhaust-
~100% recycling of
air
Total Exhaust- no
recycling of air.
Exhaust Connection
N/A
8” OD flexible duct
Exhaust CFM
20
300
Exhaust Static
N/A
Up to 2 inches w.c.
(depending upon
configuration)
Weight-
Net Weight- 525 lbs
Shipping Weight- 655 lbs
SECTION III. OPTIONAL EQUIPMENT AND ACCESSORIES-
(CONSULT YOUR SALESPERSON FOR PRICING)
LFGI-A80 Ergonomic task chair with anti-bacterial and anti-microbial vinyl covering, various
colors
LFGI-A85S Stainless steel cleaning handle with pivot head. Accepts lint- free mop heads
(LFGI-A85M)
LFGI-A8E Electrical Height Adjustment (requires additional plug-in)
LFGI-A83 Stainless steel garbage can (hangs on side to contain waste bags)
LFGI-A12 Exhaust ventilation kits
LFGI training DVD and validation documents are also available.