Operation Manual for ZBL-R620/R630 Rebar Locator 

  Press 

【↑

↓】

keys, to switch the select between “manual” and “none”; 

                  If the “manual” mode is selected, the apparatus corrects the test results by 

setting the distance from two neighboring parallel rebars to the test rebar. This function 

may improve the accuracy of the test results. The available range is 50 ~ 110 mm. To set 

the value of the distance to the neighboring rebars, first select the “manual” mode then; 

  Press

【→】

key, move the cursor to the digit which need modified; 

  Press 

【↑

↓】

keys, to amend the data; 

  Press

Save

key, save current setting, the cursor disappears, and the interface 

automatically returns to the parameter setting interface. 

Note: This function may not fully eliminate the influence coming from the 

neighboring rebars. 

7) Scan car (ZBL-R630 only) – It has two options, “on” and “off”. If the scan car is to be 

used, select “on”, otherwise select “off”. 

  Press 

Back

key, save the setting and access the testing interface.   

If it is in “parameter setting” interface 

  Press 

Menu

key, return to main menu. 

 

2.6 Scan Operation 

ZBL-R620 provides three scan modes , ZBL-R630 provides three scan modes 

1) General Test: It is the basic scan mode of R630. It accurately measures the 

thickness of concrete cover and the diameter of rebar, it may also judge if the 

component conforms to international standards. 

2) Profile Scan (ZBL-R630 only): Displays rebar’s distribution in concrete’s profile style, 

suitable for situations when rebars have sparse distribution, we suggest to use this 

sort of scan in column and plate components’ detection. 

3) Grid Scanning (ZBL-R630 only): Displays the rebar distribution over a single plane, 

suitable for situations when rebars have sparse distribution, we suggest to use this 

sort of scan in plate components’ detection. 

Introduction of shortcuts: 

 

【→】

key, adjust measuring range(status bar indicate:    / 

); 

 

【←】

key, switch save mode between “auto” and “manual” (In general test mode, 

it indicate A/M); 

 

【↑】

key, shortcut to diameter-test (suitable for general test mode); 

 

【↓】

key, switch work mode between “high-resolution rebar” and “general rebar” 

(status bar indicator: 

); 

When parameter setting is completed, press 

Back

key, the system accesses the 

“test waiting” status. 

Note: The system will reset first. When the reset is completed, it is ready to begin testing. 

The scan starting point should be selected first, and then the system will save the scan 

results. This process is discussed in more detail below. 

2.6.1 Pre-scanning 

Содержание ZBL-R620

Страница 1: ...ZBL R620 630 Rebar Locator Operation Manual...

Страница 2: ...2 2 1 System Connection 3 2 2 Startup 4 2 3 System Reset 4 2 4 Choosing Scan Mode 5 2 5 Parameter Settings 5 2 6 Scan Operation 7 2 7 Data Display 10 2 8 Data Transmission 10 2 9 Deleting Data 12 2 1...

Страница 3: ...ver and rebar diameter 1 2 Components of ZBL R620 R630 ZBL R620 R630 Rebar Locator is composed of a main instrument body a probe a scan car ZBL R630 only and signal transmission wire The probe has dir...

Страница 4: ...erface Soft keyboard interface Fig 1 2 Menu Operation Interface The testing interface is used to scan data and display image including shown in Fig 1 3 General test scan Profile scan ZBL R630 only Gri...

Страница 5: ...ig 2 1 Insert the plug black of the signal line into the socket on the left of the rebar locator rotate clockwise and lock it tightly in place Insert the other end of the signal line into the socket o...

Страница 6: ...ry must be re charged If the automatic self test is OK it automatically enters the menu display after about five seconds Shortcuts in Main Menu interface key background light on off switch keys move c...

Страница 7: ...ess highlight option Back key save current parameter setting return to previous menu and reset the system Menu key return to main menu interface Parameter amendment 1 General situation keys select the...

Страница 8: ...and the interface automatically returns to the parameter setting interface 3 Thickness setting Set the concrete cover thickness of test rebar the available range is 10 99mm the initial value is automa...

Страница 9: ...odes 1 General Test It is the basic scan mode of R630 It accurately measures the thickness of concrete cover and the diameter of rebar it may also judge if the component conforms to international stan...

Страница 10: ...d the thickness is the lowest The central line of the probe is then the accurate location of the rebar Select the central location of the two rebars which have largest distance between each other as t...

Страница 11: ...ing point unit cm will be displayed above the black dot The instrument automatically saves the test data and the image Press Back key return to parameter setting interface Press Menu key access the ma...

Страница 12: ...ce move the cursor to data display option Press Enter to access the components list interface 2 In data display interface The left column shows the components list listed in descending order with the...

Страница 13: ...ot recognized and the computer indicates that it has found new hardware device the user is directed to the appendix of R650 software operation manual for further advice 2 Computer operation Launch Reb...

Страница 14: ...plays deleting 2 10 System Settings When accessing the system setting function the information below may be amended Fig 2 16 Date Year Month Day Time Hour Minute Second 24hours mode Power status may b...

Страница 15: ...s signal cable is correct Check if the power is low battery power 0 3 3 No keypad tone Check system settings turn keyboard tones on then save settings If settings are normal but there are still no key...

Страница 16: ...ndix A General Principles of Rebar Testing Scan surface should be smooth and flat If the surface is too rough to clean put a thin board on the scan surface after completing the scan the board s thickn...

Отзывы: