Operation Manual for ZBL-R620/R630 Rebar Locator 

can work after being fully charged (or while connected to the charger). 

  If there is still no display after connecting to the charger, please check charger or 

call our company hotline. 

 

3.2 No signal 

No signal display when scanning. 

  Place the transducer away from any magnetic materials (in the air), and perform a 

system reset. Begin scan after the status bar indicates that reset is complete. 

  Check if the connection of transducer’s signal cable is correct. 

  Check if the power is low, (battery power 0%). 

 

3.3 No keypad tone 

  Check system settings, turn keyboard tones on, then save settings. 

  If settings are normal but there are still no keypad tones, call our company hotline. 

 

4. Technical Indicators 

Table 1.1: Ranges of Measuring Concrete Cover:                                                  Unit: mm 

Bar Diameter 

Concrete cover over small 

measuring range 

Concrete cover over large 

measuring range 

φ

6 ~

φ

6 ~ 70 

10 ~ 100 

φ

10 ~

φ

18 

9 ~ 80 

16 ~ 126 

φ

20 ~

φ

28 

10 ~ 86 

20 ~ 160 

φ

32 ~

φ

36 

φ

40 ~

φ

50 

16 ~ 90 

28 ~ 180 

 

Table 1.2: Measurement Accuracy (covers)                                                              Unit: mm 

Accuracy 

Concrete cover over small 

measuring range 

Concrete cover over large 

measuring range 

±

6 ~ 59 

10 ~ 79 

±

60 ~ 69 

80 ~ 119 

±

70 ~ 90 

120 ~ 180 

 

Содержание ZBL-R620

Страница 1: ...ZBL R620 630 Rebar Locator Operation Manual...

Страница 2: ...2 2 1 System Connection 3 2 2 Startup 4 2 3 System Reset 4 2 4 Choosing Scan Mode 5 2 5 Parameter Settings 5 2 6 Scan Operation 7 2 7 Data Display 10 2 8 Data Transmission 10 2 9 Deleting Data 12 2 1...

Страница 3: ...ver and rebar diameter 1 2 Components of ZBL R620 R630 ZBL R620 R630 Rebar Locator is composed of a main instrument body a probe a scan car ZBL R630 only and signal transmission wire The probe has dir...

Страница 4: ...erface Soft keyboard interface Fig 1 2 Menu Operation Interface The testing interface is used to scan data and display image including shown in Fig 1 3 General test scan Profile scan ZBL R630 only Gri...

Страница 5: ...ig 2 1 Insert the plug black of the signal line into the socket on the left of the rebar locator rotate clockwise and lock it tightly in place Insert the other end of the signal line into the socket o...

Страница 6: ...ry must be re charged If the automatic self test is OK it automatically enters the menu display after about five seconds Shortcuts in Main Menu interface key background light on off switch keys move c...

Страница 7: ...ess highlight option Back key save current parameter setting return to previous menu and reset the system Menu key return to main menu interface Parameter amendment 1 General situation keys select the...

Страница 8: ...and the interface automatically returns to the parameter setting interface 3 Thickness setting Set the concrete cover thickness of test rebar the available range is 10 99mm the initial value is automa...

Страница 9: ...odes 1 General Test It is the basic scan mode of R630 It accurately measures the thickness of concrete cover and the diameter of rebar it may also judge if the component conforms to international stan...

Страница 10: ...d the thickness is the lowest The central line of the probe is then the accurate location of the rebar Select the central location of the two rebars which have largest distance between each other as t...

Страница 11: ...ing point unit cm will be displayed above the black dot The instrument automatically saves the test data and the image Press Back key return to parameter setting interface Press Menu key access the ma...

Страница 12: ...ce move the cursor to data display option Press Enter to access the components list interface 2 In data display interface The left column shows the components list listed in descending order with the...

Страница 13: ...ot recognized and the computer indicates that it has found new hardware device the user is directed to the appendix of R650 software operation manual for further advice 2 Computer operation Launch Reb...

Страница 14: ...plays deleting 2 10 System Settings When accessing the system setting function the information below may be amended Fig 2 16 Date Year Month Day Time Hour Minute Second 24hours mode Power status may b...

Страница 15: ...s signal cable is correct Check if the power is low battery power 0 3 3 No keypad tone Check system settings turn keyboard tones on then save settings If settings are normal but there are still no key...

Страница 16: ...ndix A General Principles of Rebar Testing Scan surface should be smooth and flat If the surface is too rough to clean put a thin board on the scan surface after completing the scan the board s thickn...

Отзывы: