ELECTROMAGNETIC ENVIRONMENT GUIDANCE
INTERFERENCE
IMMUNITY TEST
IEC 60601-1-2
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT GUIDANCE
Floors should be wood, concrete,
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
The quality of the line power supply
should be that of a typical commer-
cial or hospital environment.
The quality of the line power supply
should be that of a typical commer-
cial or hospital environment.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment
The quality of the line power supply
should be that of a typical commer-
cial or hospital environment.
If the user of the Gemini laser re-
quires it to continue functioning
following interruptions of the power
supply, it is recommended to have
the Gemini laser powered by an un-
interruptible power supply or a bat-
tery.
Electrostatic
discharge (ESD)
according to IEC
61000-4-2
Electrical fast
transient/burst
according to IEC
61000-4-4
Magnetic field of
power frequen-
cies (50/60 Hz)
according to IEC
61000-4-8
Surge voltages
according to IEC
61000-4-5
Voltage dips,
short inter-
ruptions and
variations of the
power supply
according to IEC
61000-4-11
± 6 kV contact discharge
± 8 kV air discharge
± 1kV for input and
output lines
± 2 kV for power supply
lines
3 A/m
3 A/m
± 6 kV contact discharge
± 8 kV air discharge
± 1 kV differential mode
± 2 kV common mode
voltage
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
± 1 kV differential mode
± 2 kV common mode
voltage
± 1kV for input and
output lines
± 2 kV for power supply
lines
INTERFERENCE IMMUNITY
The Gemini 810 + 980 diode laser is intended for operation in the electromagnetic environment