ELECTROMAGNETIC ENVIRONMENT GUIDANCE
ELECTROMAGNETIC COMPATIBILITY
Notice: The Gemini 810 + 980 diode laser complies with all requirements for electromagnetic
compatibility according to IEC 60601-1-2: 2014.
Description: The Activation Pedal uses Bluetooth BLE 4.0 technology, which
operates at a frequency of 2402 to 2480 MHz with TX power of +0dBm and RX
sensitivity of -93dBm and uses GFSK modulation. The pedal is pre-configured
by the manufacturer to only sync with the Gemini laser unit that has a matching
unique identifier. This prevents interference with other RF wireless technologies
that may be present.
As a safety measure, any termination of the Bluetooth link between the activation
pedal and the laser unit during use will result in the immediate termination of
any laser emission. Reference the Service and Troubleshooting section of this
manual should you encounter any connectivity issues between the laser unit
and the activation pedal.
This device has passed wireless coexistence testing with common devices found
in dental practices at a minimum separation distance of 30 cm.
Accessories:
Medical grade power supply - Maximum length 6ft (1.8 meters)
Ultradent
P/N:
8981
Activation Pedal: Wireless Bluetooth at 2.4GHz
Ultradent P/N: 8982 or 8992
Medical electrical equipment needs special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the
EMC information provided in the following tables.
Portable and mobile Radio Frequency (RF) communications equipment can affect
medical electrical equipment.
The use of accessories other than those specified, except those supplied or sold by
Ultradent Products, Inc., as replacement parts for internal or external components,
may result in increased EMMISSIONS or decreased IMMUNITY of the Gemini 810 +
980 diode laser.
CAUTION
WARNING
44