WHY DOES THE LASER STOP FIRING MID-PROCEDURE OR FAIL TO TURN ON?
CAUSE:
1 - The laser diode is too hot and needs time to cool down before continued use.
2 - Output levels are more than ± 20% of the set value.
3 - Battery charge is too low to operate the laser.
SOLUTION: 1 - Allow 10-15 minutes for laser to cool down.
2 - Turn off the laser unit and wait 5 minutes, then turn the unit back on. If the laser
performs when re-tried, the microprocessor has been able to make adjustments and the
unit will function properly. If the unit fails to fire when re-tried, the device will need to
be sent for re-calibration by the manufacturer.
3 - Plug in the power adapter and let the laser unit charge for 60 minutes. You can
continue using the laser immediately after plugging in the external power.
WHY DOES THE LASER NOT FIRE WHEN I PRESS THE ACTIVATION PEDAL?
CAUSE:
1 - Activation pedal not connected.
2 - Activation pedal AA Batteries are too low to operate.
3 - Laser is in Standby mode.
SOLUTION: 1 - Check to see if the Bluetooth indicator on the activation pedal is on, and if
there is a Bluetooth indicator on the top right of the display. If either indicator is
off, re-sync the footpedal by turning the laser unit off, waiting 10 seconds, and
turning it back on. Step on the activation pedal after a wavelength mode is selected
to
re-sync.
2 - Replace the two AA batteries in the activation pedal.
3 - Touch the ACTIVE/STANDBY selection to activate the laser .
SERVICE AND TROUBLESHOOTING
I CAN HEAR THE LASER FIRING BUT IT IS CUTTING SLOWLY OR NOT CUTTING AT ALL
CAUSE:
1 - Fiber tip is not initiated.
2 - Fiber tip is not in contact with tissue.
3 - Power setting is too low.
SOLUTION: 1 - The fiber tip should be initiated for any procedures where removal of tissue is
necessary.
2 - The fiber tip should be in contact with target tissues where removal of tissue is
necessary.
3 - The laser unit should be set at the proper power setting for the procedure. The
preset procedures are a good reference point.