Chapter 1. Features and Capabilities
2
XMTC User’s Manual
1.3 Theory of Operation
The XMTC measures the concentration of a gas in a binary gas mixture by measuring the thermal conductivity of the
sample gas and comparing it to the thermal conductivity of a selected reference gas.
Two ultra-stable, glass-coated thermistors are used: one in contact with the sample gas, and the other in contact with a
selected reference gas. The thermistors are mounted so that they are in close proximity to the stainless steel walls of the
sample chamber. The entire sensor is heated to 55°C/131°F, (or 70°C/158°F) and the thermistors are heated above the
sensor temperature using a constant current source. The thermistors lose heat to the walls of the sample chamber at a
rate that is proportional to the thermal conductivity of the gas surrounding them. Thus, each thermistor will reach a
different equilibrium temperature. The temperature difference between the two thermistors is detected in an electrical
bridge circuit. It is then amplified and converted to a 4-20 mA output proportional to the concentration of one of the
constituents of the binary gas mixture. For example:
•
To measure 0 to 25% H
2
in N
2
, the reference gas would be air (2-port version, sealed reference gas), and for
calibration, the zero gas would be 100% N
2
(i.e. 0% H
2
) and the span gas would be 25% H
2
in N
2
.
•
To measure 90-100% H
2
in N
2
, the reference gas would be 100% H
2
(4-port version, flowing reference gas),
the zero gas would be 90% H
2
in N
2
, and the span gas would be 100% H
2
(the same as the reference gas).
Note:
The XMTC has polarity adjustment jumpers which permit the measurement of gases (such as CO
2
) that have a
relative thermal conductivity less than air/nitrogen.
Appendix A,
Supplemental Information
, contains a table of
Relative Thermal Conductivity of Common Gases
.
Figure 1
below shows some of these values graphically.
Figure 1: Relative Thermal Conductivity of Some Common Gases
0
1.0
2.0
3.0
4.0
5.0
6.0
7.0
SO
2
CO
2
Air/N
2
CH
4
Ne
He
H
2
C
4
H
6
Содержание XMTC
Страница 1: ...GE Measurement Control Gas Analysis 910 217 Rev G October 2013 XMTC User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii XMTC User s Manual no content intended for this page ...
Страница 18: ...Chapter 1 Features and Capabilities 8 XMTC User s Manual no content intended for this page ...
Страница 58: ...Chapter 4 Calibration 48 XMTC User s Manual no content intended for this page ...
Страница 62: ...Chapter 5 Specifications 52 XMTC User s Manual no content intended for this page ...
Страница 68: ...Appendix A Supplemental Information 58 XMTC User s Manual no content intended for this page ...
Страница 80: ...70 XMTC User s Manual no content intended for this page ...
Страница 94: ...Appendix C Installation and Wiring Diagrams XMTC User s Manual 84 no content intended for this page ...
Страница 118: ...Index 108 XMTC User s Manual no content intended for this page ...
Страница 120: ...Warranty 110 XMTC User s Manual no content intended for this page ...
Страница 124: ... no content intended for this page ...
Страница 125: ......