XMTC User’s Manual
23
Chapter 3. Operation and Programming
3.6.2 Configure Cal
The
Configure Cal
command enables you to change the field calibration type and parameters: the percentage of zero or
span gas, delay time before or after, and maximum total drift and drift of calibration. When you click on
Configure Cal
,
a window similar to
Figure 15
below opens. Clicking on any option opens the window for that option, while clicking
on
Next Item/Enter
opens the menu listed on the status line above the options.
Figure 15:
Configure Cal
Window
3.6.2a Field Cal Type
The window for
Field Cal Type
is similar to
Figure 16
below.
Figure 16:
Field Cal Type
Window
Click on
1 Point
to select 1-Point (offset, or 1-gas) calibration, or on
2-Point
to select 2-Point (zero/span, or 2-gas)
calibration. Then, click on any button on the right to return to the
Configure Cal
window.
Содержание XMTC
Страница 1: ...GE Measurement Control Gas Analysis 910 217 Rev G October 2013 XMTC User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii XMTC User s Manual no content intended for this page ...
Страница 18: ...Chapter 1 Features and Capabilities 8 XMTC User s Manual no content intended for this page ...
Страница 58: ...Chapter 4 Calibration 48 XMTC User s Manual no content intended for this page ...
Страница 62: ...Chapter 5 Specifications 52 XMTC User s Manual no content intended for this page ...
Страница 68: ...Appendix A Supplemental Information 58 XMTC User s Manual no content intended for this page ...
Страница 80: ...70 XMTC User s Manual no content intended for this page ...
Страница 94: ...Appendix C Installation and Wiring Diagrams XMTC User s Manual 84 no content intended for this page ...
Страница 118: ...Index 108 XMTC User s Manual no content intended for this page ...
Страница 120: ...Warranty 110 XMTC User s Manual no content intended for this page ...
Страница 124: ... no content intended for this page ...
Страница 125: ......