Appendix A. Supplemental Information
56
XMTC User’s Manual
A.4 Sample Calibration Sheet
A typical calibration sheet example is shown in
Figure 55
below.
Figure 55: A Sample Calibration Sheet
XMTC Calibration Sheet
XMTC S/N:
2630
XMTC Part Number:
XMTC-62-11
Calibration Part Number:
XMTC-CAL-311
Calibration Gases:
H2/N2
Calibration Ranges:
0 to 5%
Work Order Number:
508230006418
Calibration Date:
April 24, 2009
Technician:
K. Brin
Unit Calibrated Using
x10
Gain Scale
XMTC Calibration Data
Hydrogen in Nitrogen
PT
%H2
x10 (mV)
Output (mA)
1 0.00
-56.02 4.00
2 5.00
574.00 20.00
Output:
4 to 20 mA
0 to 5% H2 in N2
Clamp Output:
No
Field Calibration.
Zero:
100% N2
Span:
5% H2 in N2
Содержание XMTC
Страница 1: ...GE Measurement Control Gas Analysis 910 217 Rev G October 2013 XMTC User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii XMTC User s Manual no content intended for this page ...
Страница 18: ...Chapter 1 Features and Capabilities 8 XMTC User s Manual no content intended for this page ...
Страница 58: ...Chapter 4 Calibration 48 XMTC User s Manual no content intended for this page ...
Страница 62: ...Chapter 5 Specifications 52 XMTC User s Manual no content intended for this page ...
Страница 68: ...Appendix A Supplemental Information 58 XMTC User s Manual no content intended for this page ...
Страница 80: ...70 XMTC User s Manual no content intended for this page ...
Страница 94: ...Appendix C Installation and Wiring Diagrams XMTC User s Manual 84 no content intended for this page ...
Страница 118: ...Index 108 XMTC User s Manual no content intended for this page ...
Страница 120: ...Warranty 110 XMTC User s Manual no content intended for this page ...
Страница 124: ... no content intended for this page ...
Страница 125: ......