Appendix E. CE Mark Compliance
98
XMTC User’s Manual
E.4 Wiring the Signal Connections for the Explosion/Flameproof Version (cont.)
3.
Connect the power leads:
a.
Insert the 4-wire cable +24 VDC line (red) lead into pin TB3-1 and tighten the screw.
b.
b. Insert the 4-wire cable 24 VDC return (black) lead into pin TB3-2 and tighten the screw.
4.
Connect the analog output leads:
a.
Insert the 4-wire cable + 4-20 mA (white) lead into pin TB3-3 and tighten the screw.
b.
Insert the 4-wire cable - 4-20 mA (green) lead into pin TB3-4 and tighten the screw.
5.
Plug the EMI board back into its socket and secure it with the standoff. Terminate the ground lug of the cable to
the standoff and secure it with the provided screw.
6.
If you are installing the second cable entry device, do so in accordance with the manufacturer's instructions.
Note:
If installation of the cable entry device is only partially complete, GE recommends tagging the device to ensure
the safety of subsequent users.
a.
Remove the tab “Remove Before Installation”.
b.
Thread the cable gland entry body into the remaining XMTC port.
c.
Route the 3-wire RS232 cable & its ground lug through the cable gland.
d.
Fray the braided shield and spread evenly around the cone as shown in Figure 90 on page 97. Compress the
ring to lock the shield.
e.
After terminating the shield as shown, assemble the gland pieces together and tighten the gland to secure
the cable and the shield.
IMPORTANT:
The cable shield must be terminated in the cable gland as shown in Figure 90 on page 97.
7.
Unplug the TB2 connector by pulling it straight out of its socket, and loosen the terminal screws on the TB2
connector.
8.
Connect the RS232 serial port leads:
a.
Insert the 3-wire cable RX (red) lead into pin TB2-1 and tighten the screw.
b.
Insert the 3-wire cable TX (white) lead into pin TB2-2 and tighten the screw.
c.
Insert the 3-wire cable GND (green) lead into pin TB2-3 and tighten the screw.
9.
Carefully plug the TB2 connector back into its sockets. Terminate the ground lug to the nearest standoff.
10.
Reinstall the cover on the XMTC.
11.
Connect the other ends of the cables to the 24 VDC power supply, the 4-20 mA input of the display device, and
the serial port of the computer or terminal (see the instruction manuals for those devices for details).
CAUTION!
Connecting the +24 VDC line (red) lead to any terminal except TB3-1 will damage
the XMTC.
Содержание XMTC
Страница 1: ...GE Measurement Control Gas Analysis 910 217 Rev G October 2013 XMTC User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii XMTC User s Manual no content intended for this page ...
Страница 18: ...Chapter 1 Features and Capabilities 8 XMTC User s Manual no content intended for this page ...
Страница 58: ...Chapter 4 Calibration 48 XMTC User s Manual no content intended for this page ...
Страница 62: ...Chapter 5 Specifications 52 XMTC User s Manual no content intended for this page ...
Страница 68: ...Appendix A Supplemental Information 58 XMTC User s Manual no content intended for this page ...
Страница 80: ...70 XMTC User s Manual no content intended for this page ...
Страница 94: ...Appendix C Installation and Wiring Diagrams XMTC User s Manual 84 no content intended for this page ...
Страница 118: ...Index 108 XMTC User s Manual no content intended for this page ...
Страница 120: ...Warranty 110 XMTC User s Manual no content intended for this page ...
Страница 124: ... no content intended for this page ...
Страница 125: ......