5.11 Air Flow Control
5.11.1 Air Flow Distribution
Through the filter grid on the back of the system (Main Air Inlet), air flow into the Voluson E-Series system.
•
Air holes in the RSP power supply allow the air to pass through; the 3 fans inside the RSP suck in the
air and spread it through the beamformer.
•
By means of the 2 Backend fans, air is blown through the GEB-box (along its internal components and
the PC- Motherboard).
The warm air exits the system through holes in the left side panel (Main Air Outlet).
Figure 5-23 Main Air Inlet/Outlet at Voluson E-Series
A Main Air Inlet (filter grid)
B Main Air Outlet
Components and Functions (Theory)
Voluson E-Series Service Manual
5539550APB Revision 6
5-47
Содержание Voluson E Series
Страница 11: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 13: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 15: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 17: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 120: ...This page was intentionally left blank Setup Instructions 3 76 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 355: ......
Страница 356: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...