12. Enter the appropriate “Permanent Key” (calculated in OKOS; http://3.187.187.9/OKOS), select
OK
and
confirm with Enter.
Figure 8-5 enter Permanent key
If the entered Permanent key is correct, following window appears, confirm with Enter.
Figure 8-6 confirm key
At EC310 the format of the Permanent key has been changed.
New Format from EC310 onwards:
Previous Format still valid until EC300:
e.g.,
63XQ-4DE4-Ek1D-A3CFT-JQ0C-YSFF
e.g.,
02148EBCA0069003EF6F980F
Characters are case sensitive !!
Hyphens (-) must not be entered. They are
self-created.
Characters are not case sensitive.
13. Check disk is performed automatically - restart.
Figure 8-7 Check disk is performed automatically
14. 3 dots (one after the other) appear on the screen.
Replacement Procedures
Voluson E-Series Service Manual
5539550APB Revision 6
8-7
Содержание Voluson E Series
Страница 11: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 13: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 15: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 17: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 120: ...This page was intentionally left blank Setup Instructions 3 76 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 355: ......
Страница 356: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...