Caution
Follow the Care Card instructions supplied with each probe (inside the probe boxes) for disinfectants and
gels that are compatible with the surface material of the probes.
Note
For the latest list of compatible cleaning solutions and disinfectants refer to: http://www.gehealthcare.com/
transducers.
Please be aware of the
sensitive probe head. TAKE
EXTREME CARE!
NEVER place or store a probe
on its scan head!
When disinfecting a probe,
ensure that there is sufficient
space between the probe and
the container bottom!
10.4.5.4 Disinfection by means of the RIC-Holder
Especially for Real-time 4D endocavity probes (RIC), it is necessary to take extreme care when transporting
the system with the probe attached, or during the disinfection process. Inadequate handling may lead to
dead elements, shocked head mechanics, etc.
The RIC-Holder (especially developed for RIC Real-time 4D endocavity probes) guarantees that the
sensitive probe head does not hit the bottom of the container during the disinfection procedure.
Note
Instructions are supplied with each RIC-Holder (KTZ225469).
10.5 Using a Phantom
The use of a Phantom is not required during Preventive Maintenance. Customer may use it as part of their
Quality Assurance Program tests.
Care and Maintenance
Voluson E-Series Service Manual
5539550APB Revision 6
10-9
Содержание Voluson E Series
Страница 11: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 13: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 15: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 17: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 120: ...This page was intentionally left blank Setup Instructions 3 76 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 355: ......
Страница 356: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...