shows screw location for Mesh Shield and Penetration Panel.
NOTICE
M5 screws are used to fix Mesh Shields to the RF Shield from Magnet Room side. It’s RF
vendor’s responsibility to prepare M5 screws according to the site condition. Prepare 230
screws (and nuts). (120 pieces for ISC and 110 pieces for ICC) Mesh Shields will be installed
during System Installation by RF Vendor.
•
For Wooden RF Shield Wall: M5 Wood Screws (230 pieces)
•
For RF Shield Wall which has holes for screws: M5 screws (230 pieces) and M5 Nuts (230
pieces)
Illustration 4-18: Screw Information Of Mesh Shields (Magnet Room Side)
NOTE:
Dimensions are in inches. Bracketed dimensions are in millimeters.
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
116
4 ISC and ICC Wall Opening Requirements