Illustration 6-7: Facility Supplied System Interconnects
NOTE:
GE Healthcare recommends installing the RF Door switch on the outside wall of the
Magnet room.
The RF shielded room vendor must supply and install RF door switches on all RF
shielded doors. These switches must be wired in series and a GE supplied cable (two
loose lead conductors) will attach to one door switch. RF Switches must be rated for
10V DC maximum and the switches must be in the open position when the doors are
open (switch contacts close when the doors are completely closed).
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
Chapter 6 System Interconnections
145