7.2 Shared Equipment Rooms
When installing multiple MR systems in a shared equipment room, the following conditions must
be met:
1. The other system’s cabinet with RF amp must be separated from ISC by at least 79 in.
(2000 mm).
2. The other system’s Penetration panel must be separated from Penetration Wall of ISC by at
least 118 in (3000 mm).
3. Cables from different MR systems must not be routed together
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
Chapter 2 General System Level
41