RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
GE MEDICAL SYSTEMS
P9030GX
INSTALLATION AND REPLACEMENT
3–37
REV 21
SECTION (B) Installation of the French ROM
4.
Remove the BACK PANEL to access the MST board. Refer to illustration 3–27.
4–1. Remove the six rubber caps.
4–2. Unscrew the six screws from the back panel of the RT–x200.
4–3. Remove the back panel from the console.
SCREWS
SCREWS
BACK PANEL
BACK PANEL REMOVAL
ILLUSTRATION 3–27
5.
MST Board
5–1. Remove the MST Board (A3) from the card cage and confirm that the MST Board # is in the list below:
5–2. Remove the English ROM(s) at the location(s) listed in step 5–3.
CAUTION
Be sure the EPROM pins are straight before installation!
Be very careful to straighten pins if necessary.
5–3. Install the French ROM(s) in the following location(s) identified in the Table. Refer to illustration 3–28.
2113953
U9A
2113967–3
2113969–3
U11A
2113968–3
2113970–3
U9A
P9514NW–00
P9514QT-00
U11A
P9514NX–00
P9514QU–00
P9514ZZ–00
MST Board #
Location
English ROM Part #
French ROM Part #
P9514XJ-02
P9514XJ-03
U9A
P9514PD-01
P9514PF-01
P9514XJ-04
U9A
P9514PD-02
P9514PF-02
U11A
P9514PE-02
P9514PL-02
P9514XJ-05
U9A
P9514PD-03
P9514PF-03
U11A
P9514PE-03
P9514PL-03
P9514XH-02
U9A
P9514PT-02
P9514PV-02
U11A
P9514PU-02
P9514PW-02