RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
GE MEDICAL SYSTEMS
P9030GX
PREINSTALLATION
2–8
REV 23
2–4–2 Shipping and Storage Requirements
This unit excluding PROBES is shipped in a single container. Dimension of two types of models, namely Short and
Tall,,are given below.
TABLE 2–1
SHIPPING REQUIREMENTS
MODEL TYPE
WEIGHT kg (lbs)
Short
Monitor Container = 12.5 kg (27.5)
Console Container = 98 kg (215.6)
DIMENSION (cm)
Tall
Monitor Container = 12.5 kg (27.5)
Console Container = 100 kg (220)
49 x 45 x 43
53 x 94 x 120
49 x 45 x 43
53 x 94 x 130
Table 2–2 provides a summary of temperature, atmospheric pressure, and humidity tolerances for shipping and opera-
tion:
TABLE 2–2
STORAGE AND OPERATION REQUIREMENTS
PARAMETER
STORAGE
OPERATION
TEMPERATURE (DEG.C)
(DEG.F)
ATOMOSPHERIC PRESSURE
(hPa)
HUMIDITY (%)
(Non–condensing)
0 to 50
32 to 122
10 to 40
50 to 104
700 to 1060
700 to 1060
10 to 85
30 to 75