RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
GE MEDICAL SYSTEMS
P9030GX
INTRODUCTION
1–9
REV 23
1–2–2 Specifications
Type of protection against electric shock : Class I EQUIPMENT (System) *1
Degree of protection against electric shock : Type BF EQUIPMENT *2
Power source voltage
:US
: 115V
~
±
10%
:Europe
: 220–240V
~
±
10%
:Far East
: 100V
~
±
10%
Line Frequency : 50/60Hz (
±
2Hz)
*1.
Class 1 EQUIPMENT
EQUIPMENT in which protection against electric shock does not rely on BASIC INSULATION only, but which
includes an additional safety precaution in such a way that means are provided for the connection of ACCESSI-
BLE CONDUCTIVE PARTS to the PROTECTIVE(EARTH) CONDUCTOR in the fixed wiring of the installation in
such a way that ACCESSIBLE CONDUCTIVE PARTS cannot become LIVE in the event of a failure of the BASIC
INSULATION.
*2.
Type BF EQUIPMENT
Type B EQUIPMENT with a F-TYPE isolated applied part providing a degree of protection against electric shock
to such a degree that the allowable PATIENT LEAKAGE CURRENT under SINGLE FAULT CONDITIONS is not
exceeded when 1.1 times the highest rated MAINS VOLTAGE is applied between the APPLIED PART and earth.
– allowable LEAKAGE CURRENT;
Normal mode
Single failure mode
Patient leakage current
Less than 100
µ
A
Less than 500
µ
A