RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
GE MEDICAL SYSTEMS
P9030GX
PREINSTALLATION
2–9
REV 23
2–5 RECOMMENDED ULTRASOUND ROOM LAYOUT
Table 2–3 provides the requirements for an ultrasound room:
TABLE 2–3
ULTRASOUND ROOM REQUIREMENTS
MIN. FLOOR AREA
130 SQ FT (12 SQ M)
POWER SOURCE
CURRENT RATING
RADIATION SHIELDING
TEMPERATURE
HUMIDITY
HEAT DISSIPATION
115VAC, 60HZ or 220VAC, 50HZ, SINGLE PHASE
15A (115V) ; 10A (220V) CIRCUIT BREAKER
NONE REQUIRED for ULTRASOUND ENERGY
68–79 DEG F (20–26 DEG. C) for PATIENT COMFORT
50% to 70% for PATIENT COMFORT
650 BTU/Hr for RTfino/RT3200 Advantage–II/RT–x200
; 300 BTU/Hr per PERSON
FLOOR LOADING
about 2.4 kg/cm
2
/wheel
2–5–1 Floor Loading
The floor must be capable of supporting the RTfino/RT3200 Advantage–II/RT–x200 Console, which weighs more than
86kg(190lbs), depending on options added. The floor load is approximately 2.4kg/cm
2
/wheel.