RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
GE MEDICAL SYSTEMS
P9030GX
PERIODIC MAINTENANCE
10–4
REV 24
10–2 PERIODIC MAINTENANCE PROCEDURE (Continued)
7.
Gently clean video monitor.
CAUTION
Equipment damage possibility. Do not use a hydrocarbon glass cleaner on a monitor with an
anti-glare shield. Prolonged use of such cleaners damages the shield.
7–1. Remove the CRT (VDT) filter from the monitor.
a. Slide the left filter clamp leftwards and pull it out.
b. Slide the right filter clamp rightwards and pull it out.
c. Remove the CRT (VDT) filter.
7–2. Wipe the monitor face and CRT filter using a soft, non-abrasive folded cloth with a glass cleaner.
7–3. Install the CRT (VDT) filter after cleaning the monitor face and CRT (VDT) filter. Set the CRT (VDT) filter on
the monitor, insert the CRT filter clamps ( 1 ), and slide them inwards securely ( 2 ). Refer to ILLUSTRA-
TION 10–1.
INSTALLING CRT FILTER AND CLAMPS
ILLUSTRATION 10–1
Note
Make sure that the filter is securely fixed by the filter clamps. These are important to prevent a hazard
from the filter falling. A “click” sound will be heard when the filter clamp is secured by the lock mecha-
nism.