June 2003
2-24
Troubleshooting and Maintenance
Checking the Oxygen
Cell Calibration (cont.)
Therefore,
The new span value (x) is 100 PPMv
≅
387 µA. Enter the new value
as described in the next sub-section.
Entering the New Span
Value
1.
From the Main Menu, press the SETTINGS menu key.
2.
Enter the passcode (see page 2-6 of the
Programming Manual)
.
3.
Press the SYSTEM menu key.
4.
Press the CALIB menu key.
5.
Press the PROBE menu key until the Oxygen Probe Calibration
screen appears. See Figure 2-9 below.
Figure 2-9: System Calibration Menu for Oxygen Cell
6.
Use the CHANNEL menu key to cycle to the desired channel. The
channel number is indicated in the top right-hand-corner of the
screen. The screen will only display installed channels.
7.
Move the pointer to the Span µA line and press [YES].
8.
Enter the new span value and press [YES].
9.
Press the DONE menu key until you return to the Main Menu.
Check the screen to verify that the new span value has been
accepted.
290
100 75
–
(
)
290 0.4238
–
(
)
75 0.05
–
(
)
--------------------------------------------------------------
+
x
=
Oxygen Probe Calibration Ch
S/N
µA ppm (or %)
Zero: ____ ____
Span: ____ ____
O2 Current Multiplier:
System Calibration Menu
HELP CHANNEL PROBE DONE