June 2003
2-8
Troubleshooting and Maintenance
Procedure for Trimming
the Zero Value (cont.)
4.
Observe the multimeter reading. Wait at least 5 seconds for the
recorder output to settle. The multimeter should display one of the
following:
5.
Press the TRIM menu key. The Series 1 displays the existing
recorder trim value next to the 0% setting.The trim value will be
labeled zero or span. If the span value displays, press the ZRO/
SPAN menu key to display the zero trim value.
6.
Press the TRIM UP or TRIM DN menu key to set the trim zero
value to correct for the difference between the desired multimeter
reading and the actual multimeter reading in Step 4. Confirm that
the digital multimeter display reads the correct 0% of full scale
value (1 mA, 4 mA, or 0.1 V).
Note:
The trim resolution is limited to ±0.05 mA or ±0.5 mV. Choose
the trim value that produces an output closest to the value
desired.
7.
Press the DONE menu key to return to the Recorder Test Menu.
Note:
The zero trim is an offset adjustment, while the span trim is a
slope adjustment. As a result, the zero and span trim affect
each other. Therefore, after you adjust one, you may have to
adjust the other.
For the Recorder Output
Range
Desired Multimeter Reading
0 to 20 mA (5%)
1 mA
4 to 20 mA (0%)
4 mA
0 to 2 V (5%)
0.1 V